MRI adipose tissue segmentation and quantification in R (RAdipoSeg)

Abstract

Background

Excess adipose tissue is associated with increased cardiovascular and metabolic risk, but the volume of visceral and subcutaneous adipose tissue poses different metabolic risks. MRI with fat suppression can be used to accurately quantify adipose depots. We have developed a new semi-automatic method, RAdipoSeg, for MRI adipose tissue segmentation and quantification in the free and open source statistical software R.

Methods

MRI images were obtained from wild-type mice on high- or low-fat diet, and from 20 human subjects without clinical signs of metabolic dysfunction. For each mouse and human subject, respectively, 10 images were segmented with RAdipoSeg and with the commercially available software SliceOmatic. Jaccard difference, relative volume difference and Spearman’s rank correlation coefficients were calculated for each group. Agreement between the two methods were analysed with Bland–Altman plots.

Results

RAdipoSeg performed similarly to the commercial software. The mean Jaccard differences were 10–29% and the relative volume differences were below ( ±) 20%. Spearman’s rank correlation coefficient gave p-values below 0.05 for both mouse and human images. The Bland–Altman plots indicated some systematic and proporitional bias, which can be countered by the flexible nature of the method.

Conclusion

RAdipoSeg is a reliable and low cost method for fat segmentation in studies of mice and humans.

Background

The study of different adipose tissue depots has become highly important with the mounting evidence that visceral adipose tissue (VAT) contributes to metabolic syndrome and type 2 diabetes, while a high proportion of subcutaneous adipose tissue (SAT) is associated with a better metabolic profile [1,2,3,4]. Several methods exist for analysing the volumes of VAT, such as ultrasound [5], bioelectrical impedance analysis [6], and dual energy X-ray absorptiometry [7]. However these methods only give an indirect assessment of volume and often with poor accuracy. Computed tomography (CT) and magnetic resonance imaging (MRI) give a direct measure of VAT with high accuracy. Furthermore, these modalities enable volume analysis of other fat depots including SAT. Due to ionizing radiation exposure, CT may not always be an option in research studies. MRI does not involve ionizing radiation but has the drawback that, unlike CT, there are no fixed units. Instead, each image shows a relative difference in signal intensity between different tissues. In spite of this, MRI has shown to be as reliable as CT for fat segmentation [8].

Over the last decade, different manual, semiautomatic and automatic tools and software for fat segmentation of MRI images have been developed [9,10,11,12,13]. Concomitantly, the use and popularity of R [14] has increased greatly. R is a free and open-source software for statistical analysis and graphics. The ever-expanding library of R packages, as well as the high compatibility with other programming languages and software, makes R a highly flexible and practical tool for a wide range of tasks, including data management, statistical modelling, and visualisation. There are several packages for image manipulation and handling of medical images. However, no R protocols or packages for adipose tissue segmentation of MRI images exist.

We have integrated available packages to create a semi-automatic method for fat segmentation of MRI images in R. To test the new method, RAdipoSeg, we performed segmentation of SAT and VAT on images from lean and obese mice and humans, and compared with results from manual segmentation of the same images with the commercially available and frequently used SliceOmatic software (Tomovision, Montreal, Canada). Compared to SliceOmatic, RAdipoSeg allows for fat segmentation of MRI images with minimal cost and high flexibility regarding organisms, scanning techniques, and type of fat depots.

Methods

Tissue imaging and clinical data sampling

Images were taken from wild-type control mice in 2 different diet studies on genetically modified male mice [15, 16]. The mice were fed either a control (low-fat, LF, 3.8 kcal/g) or high-fat (HF, 4.7 kcal/g) diet for 10 weeks. Further details are presented in Additional file 3. MRI was performed in the kidney area with a 7 T small animal magnetic resonance tomograph with ParaVision 5.1 software (Bruker Biospin MRI Gmbh, Billerica, Massachusetts, USA). A series of axial base images (“base image”) were acquired with T1-weighted Rapid Acquisition with Relaxation Enhancement (RARE), and fat suppressed images (“water image”) were acquired in the same spatial domain with Chemical Shift-selective Fat Suppression (CHESS). Water suppressed images (“fat image”) were acquired by subtracting the water images from the base images. The image size was 256 × 256 voxels, with each voxel measuring 0.156 × 0.156 × 1 mm. Both MRI and glucose tolerance tests were performed in the last 3 weeks of the diet period. For the CD1 mice, glucose tolerance tests were performed by feeding the mice 2 mg/g glucose orally via tube. Blood glucose was measured at 0, 15, 30, 60, 120 and 240 min. The C57BL/6 J mice had glucose injected intraperitoneally, and blood glucose was measured at 0, 15, 30, 60, 120 and 180 min.

Human images were acquired from healthy subjects by a 2-point Dixon sequence performed on a 1.5 T Siemens Avanto running Syngo MR B17 (Syngo, Siemens, Erlangen, Germany), giving two groups of images in the same spatial domain; with the water and fat spins in-phase or out-of-phase, which is used to generate water and fat images [17]. The image size was 320 × 240 voxels, with each voxel measuring 1.234 × 1.234 × 2.5 mm.

For each mouse and each human subject, respectively, 10 sequential images in the kidney area were selected for segmentation. The kidney is located in approximately the same area in all subjects within each species, and is therefore well suited as an anchor point when comparing adipose tissue volumes. An approximate midpoint of the kidneys was found where there was an equal number of images on each side with at least 1 kidney visible. The 10 sequential images selected for volume quantification were the 5 closest images on each side of this midpoint.

R packages

This protocol was developed with R 3.6.3 in RStudio 1.2.5033 (RStudio, PBC, Boston, Massachusetts, USA). Functions are shown in cursive, and the package given in parentheses when not a basic R function. The package oro.dicom [18] was used for loading dicom files into R. Most of the functions used here are from EBImage [19] and imager [20]. The tidyverse [21] collection of packages were used for data handling and output formatting. A short description of fat segmentation is presented in Fig. 1. For a detailed protocol and further information about the packages, see Additional files 1, 2.

Fig. 1
figure 1

Overview of the workflow of fat segmentation of MRI images with RAdipoSeg. The procedure begin by finding and removing background noise, and thresholding the image. Removal of some voxels may be necessary to divide SAT from VAT, if the depots lie so close together that there is no line of black voxels between them in the image. Finally the objects are selected to the different depots and volume is calculated

Background noise and thresholding

The margins of each image were divided into 8 fields and the background noise calculated from randomly selected voxels from these fields. If some fields had high signal objects, e.g., arms, they were removed and the voxels were randomly selected from the remaining fields. The background noise (BG) was calculated as the sum of the mean and standard deviation (sd) of the voxel intensities. If objects with high signal intensity were not removed, the background noise was too high for use in further calculations.

Thresholding was used to separate objects representing adipose tissue from the rest of the image. Thresholding works by setting all voxels above or equal to a set signal intensity (the threshold) to 1 and all voxels below the threshold to 0. The threshold value was calculated from the voxel intensities along the contour of the body. The contour is the line of voxels which separates the body from the background in an MRI image. In an axial water suppressed image (fat image), a large portion of the voxels along the contour represent SAT. To find the contour, we used the base images for the mice and the in-phase images for the humans, which were thresholded using a modified version of Otsu’s method [22]. Then a contour finding algorithm was applied using ocontour (EBImage). The adipose tissue threshold for the fat images was found by subtracting sd from the mean of the intensities. When this value was below the calculated background noise, the threshold was recalculated as TC = BG + 0.1 × sd, where TC is the threshold, BG is the background noise and sd is the standard deviation of the voxel intensities, as described previously [10]. The fat images were then thresholded using thresh (EBImage). This function performs local thresholding using a moving rectangular window and TC as the offset value. The size of the rectangular window was optimized as 15 × 15 voxels for mouse images and 50 × 50 for human images.

The background noise was removed with a mask created using fillHull and floodFill (EBImage), which set all voxels outside of the mask to 0.

Labelling image objects, manual editing, and volume estimation

Some images were edited manually for removal of areas (e.g., bone marrow) and then the images were labelled using bwlabel (EBImage). This function gives each piece of interconnected voxels a separate value, counting the objects. To distinguish SAT from VAT, it was necessary to divide some objects by manually setting voxels to 0, and then relabel the image. These voxels were automatically reinserted to the appropriate depot later to avoid loss of data. Voxels can also be manually added or deleted to correct for incomplete fat suppression caused by inhomogeneties in the magnetic field during scanning. However, this was not performed on these images. The selection of objects was achieved by using grapPoint (imager). New images were made with all SAT having voxel value 1 and VAT value 2, and the volumes were calculated by counting the voxels.

Method evaluation

To evaluate the method, fat segmentation was performed on the same images with SliceOmatic’s region growing mode. Selections of the segmented images from both methods were validated by a radiologist. The volumetric overlap errors were calculated with the Jaccard distance [23], given as 100(1− (|A ∩ B|/|A  B|), where 0% is a perfect overlap between segmentation by RAdipoSeg and SliceOmatic. The relative volume differences (RVD) between RAdipoSeg (A) and SliceOmatic reference (B) were calculated as 100((|A|− |B|)/|B|). Spearman’s rank correlation coefficients were calculated to test for linear correlation. Bland–Altman analyses were performed with differences between the methods on the y-axis and mean of the methods on the x-axis, and with 95% confidence intervals. Proportional bias was tested using Spearman’s rank correlation coefficients on the data from the Bland–Altman analyses. Normalcy was assessed by Shapiro–Wilk test. All statistical analyses and making of plots were conducted in R. The figures were created in Inkscape 1.0.0 (https://inkscape.org/), with text size being the only alteration made to the plots.

Results

Mouse images were pooled into two groups of lean and obese according to diet (an overview of the groups and clinical characteristics are shown in Additional file 3). The segmentation with RAdipoSeg took approximately the same time as with SliceOmatic, although images with medium high background noise took longer to segment with RAdipoSeg than images of good quality. Images with very high levels of background noise (9 mice) were difficult to segment by SliceOmatic and were excluded from the analysis. All the human images were of good quality and hence were included (clinical characteristics of the subjects are shown in Additional file 3). Representative images for each group are shown in Fig. 2, and a summary of results and statistical analyses are presented in Table 1. The Jaccard differences were below 26% in all cases except for SAT in lean mice, and RVDs were below ( ±) 20%. Both of these measures were higher for the lean mice compared to obese mice, which is expected since the differences in a lower fat volume will influence a proportional measure more strongly than the differences in a high fat volume. The RAdipoSeg tended to give a higher volume (positive RVD) for the lean mice and a lower volume (negative RVD) for humans compared with SliceOmatic. In general, Spearman statistics had a high linear correlation between the two methods for mice and humans (Table 1 and Fig. 3). The Bland–Altman analyses (Fig. 4) showed a systematic bias for the human images (p < 0.05), but not the mouse images. There was a proportional bias of higher differences with higher volume of VAT for both humans and mice (p < 0.05). This bias persisted in Bland–Altman plots performed with proportions (SliceOmatic—RAdipoSeg/means of both methods) on the y-axis (Additional file 4). Overall, RAdipoSeg gave a volume estimation similar to SliceOmatic, but with significant systematic and proportional bias.

Fig. 2
figure 2

Comparison of fat segmentation by SliceOmatic and RAdipoSeg. One representative slice from each group of lean mice, obese mice and humans, shown from left to right as fat MRI image, SliceOmatic segmentation and RAdipoSeg segmentation. The grey background of the fat mouse MRI images show the high level of background noise

Table 1 Jaccard differences, relative volume differences and Spearman’s rank correlation coefficients
Fig. 3
figure 3

Test of linear correlation between the 2 methods for mice (n = 20) and humans (n = 20) using Spearman’s rank correlation coefficients. Volumes were calculated by summing the voxels from all images of each subject and multiplying with the voxel size. Data from the lean and obese mice were pooled together

Fig. 4
figure 4

Bland–Altman plots of the volume in cm3 of VAT and SAT for mice (n = 20) and humans (n = 20), with 1.96 ×SD limits of agreement and 95% confidence interval. Volumes were calculated by adding the voxels from all images of each subject and multiplying with the voxel size. Data from the lean and obese mice were pooled together. Spearman’s rank correlation coefficients were calculated for estimation of proportional bias

When the glucose area under the curve (AUC, blood glucose mmol/L x min) was plotted against the number of pixels representing VAT (Additional file 5), there was a slightly increasing trend for the C57BL/6 J mice, showing a higher glucose AUC with higher number of VAT pixels. The CD1 mice showed no significant difference between the diet groups in body weight, fasting glucose and glucose AUC. However, it should be noted that the CD1 mouse strain is known to be more resistant to high-fat diets than several other mice strains [24].

Discussion

In any study where adipose tissue volume is of interest, a reliable method for quantification is important. Equally important is the access to a flexible method that is available for researchers with limited funding studying any model organism. Here, we present a new method for adipose tissue segmentation and quantification, RAdipoSeg, and compare this with a frequently used, commercially available method. The 2 methods correlate well, even though the background noise in many of the mouse images caused a greater variance in this group compared to the human images.

RAdipoSeg has 2 advantages over SliceOmatic. First, SliceOmatic is an expensive software with a large annual fee in addition to a costly licence, while R is completely free. Second, RAdipoSeg sets the threshold by a standardised method, reducing subjectivity as a source of error, while the SliceOmatic threshold is set manually by the operator. This is especially important for the human images, which have a much lower range of signal intensities (below 500) compared with the mice images (around 30,000). With a lower range, a small change in the threshold will have a much larger effect. This is likely to have caused the systematic bias observed in the Bland–Altman plots of the human images; however, the bias may also have been caused by the automatic threshold set by RAdipoSeg being too high, rather than the manual threshold of SliceOmatic being set too low. A study with a larger sample size, which also looks into the inter- and intra-reproducibility of both methods, may answer this question.

For VAT we observed a proportionial bias for both human and mouse images. With a larger tissue volume there is also a larger region of interest, which increases the number of voxels that can be mislabelled. We would therefore expect more differences the larger the volume. However, when plotting proportional differences against the means the bias did not disappear. We also observed a higher Jaccard difference and RVD for VAT compared with SAT in humans. This and the proportional bias may be caused by the offset value used for thresholding with RAdipoSeg being based on the signal intensity of SAT. When there is a higher variation in the signal intensity of VAT compared with SAT, the adipose voxels with low signal in VAT could be excluded from segmentation. The selection of human subjects in this study was limited and consisted of metabolically healthy controls. In a study with more variation in visceral adipose tissue between the human subjects, the segmentation of individuals with a higher VAT volume may give an underestimated quantification of VAT. This could in turn underestimate the differences between the subjects in the study, and create a bias in the data towards a lower VAT volume. If the automatically generated threshold is seen as inadequate, only a minor alteration of RAdipoSeg is required for setting a manual threshold. This would prevent the underestimation of volume in subjects with a high level of adipose tissue, at the cost of an increase in subjectivity. In many studies, having a standardised method might be preferable to gaining a more exact adipose tissue volume estimation.

Image quality was a factor in comparing the two methods. The CHESS technique of fat suppression is known to be sensitive to B0 and B1 heterogeneity, which may be the reason for the incomplete fat suppression observed in many of the mouse images. It was difficult to segment images with high background noise using SliceOmatic. The threshold had to be set very low and the fat had to be subjectively separated from the background noise. RAdipoSeg gave good segmentation of these images, but lower image quality increased the duration of the procedure by requiring more manual interference. RAdipoSeg uses a local threshold, which allows for better separation between signal and background noise than a global threshold. Each voxel is analysed based on the neighbouring voxels, while global threshold is affected by artefacts in the entire image. Also, with RAdipoSeg, the noise outside of the region of interest is automatically deleted by the mask, which gives a clear boundary against the background. With SliceOmatic the boundary had to be decided manually during segmentation. In addition to the low sample size, this may be the reason for the poor Spearman’s correlation coefficient between the methods for SAT in obese mice. Image quality was not a factor for the human images, as they all had little background noise.

Only the region growing module of the SliceOmatic software could be used in this study, as the high level of background noise made the other modules impractical or impossible to use. For the human images, the other modules for fat segmentation offered by SliceOmatic were available and may have yielded a different result.

There are several automatic and deep learning software for fat segmentation of MRI images, such as FatSegNet [12] and an algorithm for MatLab (MathWorks, Natick, Massachusetts) [13]. However, most of these are specialized for human images and, in some cases, for specific human regions. Others require a high number of images for training. For smaller studies on images from rodents, an automatic method designed for human images may not be easily adapted. In comparison, RAdipoSeg is flexible and can be used on any MRI images from any organism. By implementing other programming languages, e.g., Python, RAdipoSeg can be further automatised, thereby improving speed and usability of the procedure (Additional files 6 and 7).

There was a slightly increasing trend for the correlation between VAT volume and blood glucose AUC for the C57BL/6 J mice. Since ectopic accumulation of lipids in the liver is an important predictor for glucose intolerance [25], it should be noted that RAdipoSeg may also be adapted to measure liver adiposity. Liver percentage fat fraction can be measured using the method Iterative decomposition of water and fat with echo asymmetry and least squares estimation (IDEAL) [26]. RAdipoSeg may be able to pick out the geographical position of the liver from MRI images where the liver, or the immediate surrounding tissues, have a high pixel intensity. The geographical location can then be used to measure percentage liver fat fraction from the corresponding IDEAL MRI images.

Conclusions

We present and validate a novel method for fat segmentation of MRI images, RAdipoSeg, in the free and open-source software R. A comparison with the SliceOmatic software shows that RAdipoSeg can be used to give a reliable and consistent volume estimate of fat in studies of mice and humans. Though a proportional bias was detected for images with a high adipose tissue volume, this can be countered by setting the threshold manually. Furthermore, though the fat suppression technique will affect the results to some degree, this new method can be used on images acquired either by the CHESS or the 2-point Dixon technique. RAdipoSeg is therefore a suitable and reliable method for relative comparisons of fat depot images from studies with standardised imaging.

Availability of data and materials

All data generated and analysed during this study are included in this published article and its supplementary information files. The MRI images analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AUC:

Area under the curve

BG:

Background noise

BMI:

Body mass index

BW:

Body weight

CHESS:

Chemical shift-selective fat suppression

HF:

High-fat diet

IDEAL:

Iterative decomposition of water and fat with echo asymmetry and least squares estimation

LF:

Low-fat control diet

RARE:

Rapid acquisition with relaxation enhancement

REK:

Western norway regional committee for medical research ethics

RVD:

Relative volume difference

SAT:

Subcutaneous adipose tissue

TC:

Threshold

VAT:

Visceral adipose tissue

References

  1. Marin P, Andersson B, Ottosson M, Olbe L, Chowdhury B, Kvist H, et al. The morphology and metabolism of intraabdominal adipose tissue in men. Metabolism. 1992;41(11):1242–8.

    Article  CAS  Google Scholar 

  2. Brochu M, Tchernof A, Dionne IJ, Sites CK, Eltabbakh GH, Sims EA, et al. What are the physical characteristics associated with a normal metabolic profile despite a high level of obesity in postmenopausal women? J Clin Endocrinol Metab. 2001;86(3):1020–5.

    CAS  PubMed  Google Scholar 

  3. Matsha TE, Ismail S, Speelman A, Hon GM, Davids S, Erasmus RT, et al. Visceral and subcutaneous adipose tissue association with metabolic syndrome and its components in a South African population. Clin Nutr ESPEN. 2019;32:76–81.

    Article  Google Scholar 

  4. Shah RV, Murthy VL, Abbasi SA, Blankstein R, Kwong RY, Goldfine AB, et al. Visceral adiposity and the risk of metabolic syndrome across body mass index: the MESA study. JACC Cardiovasc Imaging. 2014;7(12):1221–35.

    Article  Google Scholar 

  5. Suzuki R, Watanabe S, Hirai Y, Akiyama K, Nishide T, Matsushima Y, et al. Abdominal wall fat index, estimated by ultrasonography, for assessment of the ratio of visceral fat to subcutaneous fat in the abdomen. Am J Med. 1993;95(3):309–14.

    Article  CAS  Google Scholar 

  6. Nagai M, Komiya H, Mori Y, Ohta T, Kasahara Y, Ikeda Y. Development of a new method for estimating visceral fat area with multi-frequency bioelectrical impedance. Tohoku J Exp Med. 2008;214(2):105–12.

    Article  Google Scholar 

  7. Treuth MS, Hunter GR, Kekes-Szabo T. Estimating intraabdominal adipose tissue in women by dual-energy X-ray absorptiometry. Am J Clin Nutr. 1995;62(3):527–32.

    Article  CAS  Google Scholar 

  8. Seidell JC, Bakker CJ, van der Kooy K. Imaging techniques for measuring adipose-tissue distribution–a comparison between computed tomography and 1.5-T magnetic resonance. Am J Clin Nutr. 1990;51(6):953–7.

    Article  CAS  Google Scholar 

  9. Tang Y, Lee S, Nelson MD Jr, Richard S, Moats RA. Adipose segmentation in small animals at 7T: a preliminary study. BMC Genomics. 2010;11(Suppl 3):S9.

    PubMed  PubMed Central  Google Scholar 

  10. Thomas EL, Saeed N, Hajnal JV, Brynes A, Goldstone AP, Frost G, et al. Magnetic resonance imaging of total body fat. J Appl Physiol. 1998;85(5):1778–85.

    Article  CAS  Google Scholar 

  11. Leporq B, Lambert SA, Ronot M, Vilgrain V, Van Beers BE. Quantification of the triglyceride fatty acid composition with 3.0 T MRI. NMR Biomed. 2014;27(10):1211–21.

    Article  CAS  Google Scholar 

  12. Estrada S, Lu R, Conjeti S, Orozco-Ruiz X, Panos-Willuhn J, Breteler MMB, et al. FatSegNet: a fully automated deep learning pipeline for adipose tissue segmentation on abdominal dixon MRI. Magn Reson Med. 2020;83(4):1471–83.

    Article  Google Scholar 

  13. Sadananthan SA, Prakash B, Leow MK, Khoo CM, Chou H, Venkataraman K, et al. Automated segmentation of visceral and subcutaneous (deep and superficial) adipose tissues in normal and overweight men. J Magn Reson Imaging. 2015;41(4):924–34.

    Article  Google Scholar 

  14. R Core Team. R A language and environment for statistical computing. Vienna: R Foundation for Statistical Computing; 2019.

    Google Scholar 

  15. Bjune JI, Haugen C, Gudbrandsen O, Nordbo OP, Nielsen HJ, Vage V, et al. IRX5 regulates adipocyte amyloid precursor protein and mitochondrial respiration in obesity. Int J Obes (Lond). 2019;43(11):2151–62.

    Article  CAS  Google Scholar 

  16. Svard J, Rost TH, Sommervoll CEN, Haugen C, Gudbrandsen OA, Mellgren AE, et al. Absence of the proteoglycan decorin reduces glucose tolerance in overfed male mice. Sci Rep. 2019;9(1):4614.

    Article  Google Scholar 

  17. Tjora E, Wathle G, Engjom T, Erchinger F, Molven A, Aksnes L, et al. Severe pancreatic dysfunction but compensated nutritional status in monogenic pancreatic disease caused by carboxyl-ester lipase mutations. Pancreas. 2013;42(7):1078–84.

    Article  CAS  Google Scholar 

  18. Whitcher B, Schmid VJ, Thornton A. Working with the DICOM and NIfTI data standards in R. J Stat Softw. 2011;44(6):1–28.

    Article  Google Scholar 

  19. Pau G, Fuchs F, Sklyar O, Boutros M, Huber W. EBImage–an R package for image processing with applications to cellular phenotypes. Bioinformatics. 2010;26(7):979–81.

    Article  CAS  Google Scholar 

  20. Barthelme S. Image processing library based on CImg. J Open Source Softw. 2019;4:38. https://doi.org/10.2110/joss.01012.

    Article  Google Scholar 

  21. Wickham H, Averick M, Bryan J, Chang W, McGowan LD, François R, et al. Welcome to the tidyverse. J Open Source Softw. 2019;4(43):1686.

    Article  Google Scholar 

  22. Otsu N. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern. 1979;9(1):62–6.

    Article  Google Scholar 

  23. Jaccard P. Etude de la distribution florale dans une portion des alpes et du Jura. Bull Soc Vaud Sci Nat. 1901;37(142):547–79.

    Google Scholar 

  24. Fengler VH, Macheiner T, Kessler SM, Czepukojc B, Gemperlein K, Muller R, et al. Susceptibility of different mouse wild type strains to develop diet-induced NAFLD/AFLD-associated liver disease. PLoS ONE. 2016;11(5): e0155163.

    Article  Google Scholar 

  25. Shulman GI. Ectopic fat in insulin resistance, dyslipidemia, and cardiometabolic disease. N Engl J Med. 2014;371(12):1131–41.

    Article  Google Scholar 

  26. Procter AJ, Sun JY, Malcolm PN, Toms AP. Measuring liver fat fraction with complex-based chemical shift MRI: the effect of simplified sampling protocols on accuracy. BMC Med Imaging. 2019;19(1):14.

    Article  Google Scholar 

Download references

Acknowledgements

The magnetic resonance imaging of the mice was performed at the Molecular Imaging Center (MIC), Department of Biomedicine, University of Bergen. We wish to extend our special thanks to Carol Cook for her valuable contributions to this study.

Funding

Open access funding provided by University of Bergen. This study was funded by Western Norway Regional Health Authority (https://helse-vest.no/en) and Trond Mohn Foundation (https://mohnfoundation.no/). The funders had not and will not have a role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

CH created the code, designed the experiment and drafted the manuscript. VL added to and edited the code. IH validated a selection of segmented images, and IH and ET provided the human MRI images. CH and OAG provided the mouse images. OAG, JVS, SND and GM supervised the study. All authors edited the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Gunnar Mellgren.

Ethics declarations

Ethics approval and consent to participate

All animal studies were approved by the Norwegian State Board of Biological Experiments with Living Animals and carried out in accordance with their guidelines. The acquisition and further use of the anonymised images were approved by the Western Norway Regional Committee for Medical Research Ethics (REK). The acquisition of the images was performed according to REK’s guidelines and each of the subjects gave written informed consent.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

: can be viewed in RStudio (https://www.rstudio.com/) Complete code for performing adipose tissue segmentation with RAdiposeg. This format allows for convenient and direct use of the code in RStudio.

Additional file 2.

Complete code for performing adipose tissue segmentation with RAdiposeg, read format. This format allows for reading the entire code without RStudio.

Additional file 3.

Characteristics of the mouse and human studies/groups used for the fat segmentation.

Additional file 4.

Bland-Altman plots with proportions on the y-axes. Plots calculated with volume in cm3 of VAT and SAT for mice (n=20) and humans (n=20), with 1.96 ×SD limits of agreement and 95 % confidence interval. Volumes were calculated by adding the voxels from all images of each subject and multiplying with the voxel size. Data from the lean and obese mice were pooled together. Spearman’s rank correlation coefficients were calculated for estimation of proportional bias.

Additional file 5.

Blood glucose AUC plotted against VAT volume. The plot was separated by methods, RAdipoSeg and Sliceomatic, and mouse strain, CD1 and C57BL/6J. Test of linear correlations were performed using Spearman’s rank correlation coefficients. P-values (rho) for the CD1 mice were 0.84 (− 0.10) for RAdipoSeg and 0.93 (0.05) for SliceOmatic. P-values (rho) for the C57BL/6J mice were 0.01 (0.79) for RAdipoSeg and 0.01 (0.78) for SliceOmatic.

Additional file 6.

MRI image data. All data collected from each MRI image using the RAdipoSeg and SliceOmatic segmentation methods.

Additional file 7.

Readme file with details about Additional file 6.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Haugen, C., Lysne, V., Haldorsen, I. et al. MRI adipose tissue segmentation and quantification in R (RAdipoSeg). Diabetol Metab Syndr 14, 146 (2022). https://doi.org/10.1186/s13098-022-00913-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13098-022-00913-x

Keywords