Systematic review and meta-analysis of the effectiveness of continuous glucose monitoring (CGM) on glucose control in diabetes

Abstract

Diabetes mellitus is a chronic disease that necessitates continuing treatment and patient self-care education. Monitoring of blood glucose to near normal level without hypoglycemia becomes a challenge in the management of diabetes. Although self monitoring of blood glucose (SMBG) can provide daily monitoring of blood glucose level and help to adjust therapy, it cannot detect hypoglycemic unawareness and nocturnal hypoglycemia which occurred mostly in T1DM pediatrics. Continuous glucose monitoring (CGM) offers continuous glucose data every 5 minutes to adjust insulin therapy especially for T1DM patients and to monitor lifestyle intervention especially for T2DM patients by care providers or even patients themselves. The main objective of this study was to assess the effects of continuous glucose monitoring (CGM) on glycemic control in Type 1 diabetic pediatrics and Type 2 diabetic adults by collecting randomized controlled trials from MEDLINE (pubmed), SCOPUS, CINAHL, Web of Science and The Cochrane Library up to May 2013 and historical search through the reference lists of relevant articles. There are two types of CGM device: real-time CGM and retrospective CGM and both types of the device were included in the analysis. In T1DM pediatrics, CGM use was no more effective than SMBG in reducing HbA1c [mean difference – 0.13% (95% CI -0.38% to 0.11%,]. This effect was independent of HbA1c level at baseline. Subgroup analysis indicated that retrospective CGM was not superior to SMBG [mean difference -0.05% (95% CI -0.46% to 0.35%)]. In contrast, real-time CGM revealed better effect in lowering HbA1c level compared with SMBG [mean difference -0.18% (95% CI -0.35% to -0.02%, p = 0.02)]. In T2DM adults, significant reduction in HbA1c level was detected with CGM compared with SMBG [mean difference – 0.31% (95% CI -0.6% to -0.02%, p = 0.04)].

This systematic review and meta-analysis suggested that real-time CGM can be more effective than SMBG in T1DM pediatrics, though retrospective CGM was not. CGM provided better glycemic control in T2DM adults compared with SMBG.

Introduction

The global prevalence of diabetes by International Diabetes Federation (IDF) estimation shows that there are 366 million people with diabetes in 2011, and this is expected to rise to 552 million by 2030 [1]. Diabetes caused 4.6 million deaths in 2011 [2]. Most of people afflicted with diabetes (85 – 95%) have type 2 diabetes mellitus [3]. Forty-four percent of people with type 2 diabetes show hemoglobin A1C higher than the generally accepted target [4]. Recent studies have established that improved glycemic control by means of intensive diabetes management decreases the risk of long-term microvascular complications in both type 1 and type 2 diabetic patients [5]. However, the major problem with an intensive diabetes treatment is the increased risk of hypoglycemic events. Subjects in the intensive management group of the Diabetes Control and Complications Trial (DCCT) experienced severe hypoglycemia two to three times more often than subjects receiving conventional therapy [6]. The fear of hypoglycemia often leads patients to forget the fatal consequences of long-term complications resulting in loss of control and cognitive dysfunction [7]. Therefore, in addition to intensive treatment, self management of blood glucose to normal or within normal limits is a crucial aspect.

Self monitoring of blood glucose (SMBG) has been shown to be as effective in insulin-treated type 1 and type 2 diabetes. Although the effect of SMBG already demonstrated in some meta-analysis [8, 9], it is not recommended as regularly use in non-insulin treated type 2 diabetes. SMBG fails to detect nocturnal hypoglycemia and asymptomatic hypoglycemia even in patients with good control of HbA1c values and it needs multiple blood samples throughout the day. In addition, SMBG gives a single instant reading without any information on glucose trends and thus may miss important and significant glucose fluctuations [1012]. Continuous glucose monitoring (CGM) measures interstitial fluid every 10 seconds and an average glucose value is recorded every five minutes 24 hours a day. This gives a more accurate pattern of daily glucose fluctuations allowing identification of the glycemic effect of food, physical activity, insulin and different medication types and doses aiding in better self management with avoiding unrecognized hypoglycemia [13].

Although some randomized trials show the benefit of CGM use over SMBG in type 1 diabetic pediatric population [14, 15], previous meta-analysis on type 1 diabetic pediatric patients failed to show the effectiveness of CGM on the decline of HbA1c level compared with SMBG [16], further evidence is still needed to provide effectiveness of CGM on pediatric type 1 diabetic population. Only one meta-analysis reported that the effectiveness of CGM on type 2 diabetic patients was superior to SMBG use [17]. However, the strength of evidence of reducing HbA1c in that population is limited because the number of patients in the included studies was rather low. In our meta-analysis, the effectiveness of CGM use over SMBG was evaluated in Type 1 diabetic pediatrics and Type 2 diabetic adults to provide a comprehensive and quantitative synthesis of evidence from all randomized controlled trials (RCTs) and to make clear the conflict about CGM effectiveness in these populations.

Methods

Data Sources

Reports of randomized controlled trials of CGM aimed for good glycemic control in type 1 diabetic and type 2 diabetic patients are identified through a systematic literature search of MEDLINE (Pubmed), Scopus, CINAHL, Web of Science and The Cochrane Library. The bibliographic databases were searched from the inception to May 2013. The following MeSH terms were used; diabetes mellitus and continuous glucose monitoring. This was followed by keyword search using as keywords continuous monitoring of blood glucose, CGM, insulin-dependent diabetes mellitus, IDDM, non-insulin dependent diabetes mellitus and NIDDM. Historical search of reference lists of relevant randomized controlled trials, systematic and narrative reviews was also undertaken. No language restriction was imposed.

Inclusion criteria and exclusion criteria

To be included in the systematic review, studies had to be

  1. a)

    randomized controlled trials comparing currently available CGM device with SMBG in patients with type 1 diabetes pediatrics (≤ 18 yrs) or type 2 diabetes adults (≥ 18 yrs),

  2. b)

    of at least 8 weeks duration, and

  3. c)

    reporting HbA1c as an outcome measure.

We excluded studies involving

  1. a)

    pregnant women,

  2. b)

    critically ill patients, post-surgery, post-transplant and ICU patients.

Data extraction and study quality assessment

Data from individual studies were abstracted and study quality was evaluated independently by two authors using a standardized form. The disagreements were resolved by a third author. The methodological quality of each study was assessed using Maastricht Amsterdam scale [18], which has been developed based on the scale of Jadad et al [19] and the Delphi list [20]. These 12 items evaluated the internal validity of the study results. Each item had a rating scale of “yes,” “no,” or “unsure.” If bias was unlikely, the item was rated positive. If bias was likely, the item was rated as negative. Then, if information concerning the item was not available, it was rated with “unsure.” Out of 12 items, only 11 items were used to assess the quality of studies because one item that investigates patient compliance is not applicable to CGM intervention. Studies that met at least 6 of 11 quality criteria were of high quality. Those scoring less than 6 of the criteria were of low quality or having high risk of bias.

Statistical analysis

Outcome of interest was HbA1c. Treatment effect was estimated with mean difference in the final values of HbA1c between the CGM group and the SMBG group. In a randomized trial, a comparison of final measurement can usually be assumed to provide the same estimate as the comparison of changes from baseline. Random effects model was used to combine the results of individual studies when Q-statistic test was significant at the level of 0.1, otherwise the fixed effects model was used. Heterogeneity was also determined by I2. Substantial heterogeneity was represented by I2 of 50% or more [21]. Subgroup analysis was carried out based on the data presentation of CGM device [real-time (RT) CGM and retrospective (r) CGM] and study quality. Moreover, subgroup analysis based on baseline HbA1c level (< 8%, 8-10%, and >10%) was also performed to assess the impact of baseline HbA1c level on the effectiveness of CGM. Sensitivity analysis was conducted by excluding studies reporting lower number of usable CGM data. The statistical analysis was undertaken with Review Manager (Revman®) program version 5.2.3 (Cochrance collaboration, Oxford UK). A funnel plot and Egger regression test [22] were used to assess publication bias.

Results

Study characteristics

In pediatric Type 1 diabetes, ten RCTs involving 817 participants met our predefined inclusion criteria. The diagram of data extraction is illustrated in Figure 1. The characteristics of included trials are summarized in Table 1. Seven studies are regarded as high quality [14, 15, 2327] and three studies are regarded as low quality [2830]. Retrospective CGM (r-CGM) was used in 5 studies [2325, 28, 29] and Real-Time CGM (RT-CGM) was used in 5 studies [14, 15, 26, 27, 30]. Duration of study of all trials ranged from 3 months to 12 months. Five studies [2426, 28, 30] included participants with insulin pump therapy or insulin injection therapy, three studies [14, 23, 29] included only participants with insulin injection therapy and two studies [15, 27] included only participants with insulin pump therapy. Among ten studies, four studies were multi-centered [14, 15, 26, 27]. There were two single-blinded studies [24, 30] and one double-blinded study [23]. Two out of ten studies were cross-over randomized trials [15, 23].

Figure 1
figure 1

Summary of trial flow (Type 1 diabetic pediatrics).

Table 1 Characteristics of included studies in Type 1 diabetic pediatrics

In Type 2 diabetes, five RCTs involving 161 participants met our predefined inclusion criteria [3135]. The diagram of data extraction is illustrated in Figure 2. Among these 5 trials, there was a study which analyzed separately for two follow-up periods; i.e, 26 weeks and 52 weeks, and presented the results in two separated papers [34, 35]. Only the paper with intervention period was included in the analysis [34]. Therefore, only 4 randomized trials were included in meta-analysis. The characteristics of included trials are summarized in Table 2. Two studies were of high quality [31, 33] and other two studies were of low quality [32, 34]. Among 4 trials, only one trial is multi-centered study [32]. One study was conducted to test the effects of a counseling intervention using continuous glucose monitoring system feedback on physical activity self-efficacy and reported HbA1c as a secondary outcome [33]. Two studies were performed in T2DM patients who were not treated with insulin [33, 34] and other two included participants treated with either oral hypoglycemic agents (OHA) or insulin injections [31, 32].

Figure 2
figure 2

Summary of trial flow (Type 2 diabetic adults).

Table 2 Characteristics of included studies in Type 2 diabetic adults

Type 1 diabetic pediatrics

Total analysis

There were a total of 817 type 1 diabetic pediatrics (413 in CGM group and 404 in SMBG group) in ten trials that compared CGM use with SMBG. Significant heterogeneity was found among the study results. The reason may be because of variations in patients’ characteristics, intervention used, pattern and frequency of CGM use, and quality of studies. The use of CGM did not provide better efficacy than SMBG use in terms of HbA1c reduction in T1DM pediatrics. The pooled estimate of HbA1c difference was – 0.13% (95%CI -0.38% to 0.11%, p = 0.27) (Figure 3). Publication bias was detected (Egger: bias = -3.19 (95% CI = -6.35 to -.028) P = 0.048 (Figure 4).

Figure 3
figure 3

Mean difference (95% confidence interval) in HbA1c for CGM versus SMBG in T1DM pediatrics.

Figure 4
figure 4

Funnel plot for ten randomized controlled trials of CGM versus SMBG in T1DM pediatrics.

Subgroup analysis based on type of data presentation

Retrospective CGM vs. SMBG

Retrospective CGM (r-CGM) was used in 5 trials [2325, 28, 29]. The r-CGM was no better than SMBG use for glycemic control in T1DM pediatrics. The pooled mean difference in HbA1c was -0.05% (95% CI -0.46% to 0.35%, p = 0.79) (Figure 5).

Figure 5
figure 5

Mean difference (95% confidence interval) in HbA1c for retrospective CGM versus SMBG in T1DM pediatrics.

Real-time CGM vs. SMBG

Real time CGM (RT-CGM) was used in 5 trials [14, 15, 26, 27, 30]. The use of RT-CGM was superior to the usual care in improving glycemic control. The pooled mean difference in HbA1c was -0.18% (95% CI -0.35% to -0.02%, p = 0.02) (Figure 6).

Figure 6
figure 6

Mean difference (95% confidence interval) in HbA1c for real-time CGM versus SMBG in T1DM pediatrics.

Subgroup analysis based on HbA1c at baseline

Subgroup analysis was also performed according to baseline HbA1c of participants in the studies (HbA1c < 8%, HbA1c 8-10%, HbA1c > 10%). There were 2 studies with HbA1c at baseline < 8%. The results suggested that CGM was no more effective than SMBG in glycemic control regardless of baseline HbA1c level. The pooled mean difference in HbA1c was -0.01% (95%CI -0.23% to 0.20%, p = 0.91), -0.19% (95%CI -0.61% to 0.23%, p = 0.38) and -0.23% (95%CI -0.58% to 0.13%, p = 0.21), respectively, for baseline HbA1c < 8%, 8-10%, and > 10% (Figure 7).

Figure 7
figure 7

Mean difference (95% confidence interval) in HbA1c for CGM versus SMBG according to HbA1c at baseline in T1DM pediatrics.

Subgroup analysis based on quality of included studies

Subgroup analysis was also performed according to quality of studies (high quality studies and low quality studies). There were 7 high quality studies and 3 low quality studies. The pooled mean difference in HbA1c was -0.25% (95%CI -0.57% to 0.07%, p = 0.12) for high quality studies and 0.18% (95%CI -0.16% to 0.52%, p = 0.30) for studies with low quality (Figure 8).

Figure 8
figure 8

Mean difference (95% confidence interval) in HbA1c for CGM versus SMBG according to quality of included studies in T1DM pediatrics.

Sensitivity analysis

Sensitivity analysis was performed by excluding two studies which described lower number of usable CGM data [25, 28]. For example, one study reported that only 70% of the total hours worn was usable and this suboptimal data was because of insufficient calibration points [25], while the other described that the lower numbers of usable CGM data was available because of patient noncompliance [28]. Participants in Chase et al. study [28] attained an average of less than 50% (421 of 864 possible) usable readings per sensor. We have doubted that whether this percentage of data would represent the overall effect on experimental population. Enough explanation for usable data was not provided in individual study and there are no criteria to identify the amount of the CGM data to represent the experimental group. Therefore, this may bias the result in favor of SMBG. Heterogeneity was reduced from 71% to 35% when these two trials were excluded. Glycemic control was better with CGM than SMBG (mean difference -0.23%, 95% CI -0.38% to -0.08%, p = 0.003) (Figure not shown).

Type 2 diabetic adults

There were a total of 228 type 2 diabetic adults (111 in the intervention group and 117 in the control group) in four trials that compared CGM use with SMBG. The use of CGM provided better efficacy than SMBG use in terms of HbA1c reduction in T2DM adults. The pooled mean difference in HbA1c was – 0.31% (95%CI -0.6% to -0.02%, p = 0.04) (Figure 9). No publication bias was detected (Egger: bias = 0.29, 95% CI = -1.77 to 2.36) (Figure not shown).

Figure 9
figure 9

Mean difference (95% confidence interval) in HbA1c for CGM versus SMBG in T2DM adults.

Discussion

Type 1 diabetic pediatrics

According to American Academy of Pediatrics, the term ‘pediatric’ includes population from birth to 18 years old [36]. In our meta-analysis, we then used the term ‘pediatric’ and included studies with T1DM participant ≤ 18 years of age. The principal meta-analysis evaluated the effect of both retrospective CGM and real-time CGM altogether and included 10 randomized controlled trials. The pooled results revealed that as a whole CGM was no more effective than SMBG in lowering HbA1c [mean difference -0.13%, 95%CI -0.38% to 0.11%). The results support the conclusion previously reported by Golicki et al (mean difference -0.02%, 95%CI -0.29 to 0.25) [16]. Their meta-analysis of CGM in children included five studies from the age of 2 to 19 years [16]. Our meta-analysis included ten studies, four of which already included in their meta-analysis. One study that was included in the previous meta-analysis was excluded from ours because participants were older than 18 years [37]. Although study participants aged < 18 years were eligible, those aged 2-18 years were enrolled among the trials contributed to our meta-analysis since there was no randomized study on participants with age ≤ 2 years. In addition, their meta-analysis limited to only one device, Metronic MiniMed. In contrast, our meta-analysis included all kinds of CGM devices which are available in the markets in the USA or Europe since the purpose of our study was to investigate a treatment strategy using CGM technology and not a specific device.

Inconsistency among studies as assessed by I2 statistic was statistically significant (I2 = 71%). There was difference in sample size among studies ranging from 11 to 156 participants. Other possible reasons for heterogeneity may be the differences in frequency and duration of CGM use, intervention period, and intervention used among those studies (retrospective or real-time CGM studies). In all studies with retrospective CGM, the device was used for 72 consecutive hours but with different frequency for different duration, for example, 6 times in 1 month [28], 1 time in 3 months [29], 4 times in 3 months [25], 3 times in 4 months [24], and 1 time in 3 months [23]. For RT-CGM, the device was used on a daily basis except in the study by Bergenstal et al. [14] in which the device was used for 1-week period continuously per 6 months. Subgroup analysis based on the frequency of CGM use was not possible. In addition, various types of insulin delivery systems were employed among individual trials, for example, both insulin pump and insulin injection therapies [2426, 28, 30], insulin injection alone [14, 23, 29], and insulin pump therapy alone [15, 17].

It has been demonstrated that patients who used CGM devices had a greater decrease in hemoglobin A1c from baseline compared with those using SMBG in type 1 diabetics patients [mean difference – 0.26% (95%CI -0.34% to -0.19%)]. However, when the effectiveness of the r-CGM and RT-CGM compared with SMBG were analyzed separately, only RT-devices for CGM improved glycemic control [mean difference -0.27% (95%CI-0.34% to -0.19%)] [38]. The beneficial effect of RT- CGM over SMBG was consistently described in type 1 pediatric population [39]. Other meta-analysis of CGM in pediatric population also reported no significant effect of r- CGM compared with SMBG [16]. Therefore, although RT-CGM gave profound benefit to type 1 diabetic population, the value of r-CGM was somewhat limited. These results are subsequently confirmed by our meta-analysis, showing that as a whole CGM did not provide better benefit over SMBG. Specifically, glycemic control was better with RT-time CGM compared with SMBG [mean difference -0.18% (95% CI -0.35% to -0.02%, p = 0.02)], whereas it was no better with r-CGM than with SMBG [mean difference -0.05% (95% CI -0.46% to 0.35%)]. As a matter of fact, regarding patient perspective, learning about the application of device can be a burden to patients in using RT-CGM. R-CGM seems to be more convenient since patients need doing nothing. However, RT-CGM proved better glycemic control than r-CGM. For r-CGM, physicians or care providers can adjust therapy based on CGM data. For RT-CGM, physicians make therapeutic adjustment based on CGM data and, at the same time, patients can also monitor lifestyle intervention themselves. In other words, RT-CGM can be a supportive tool for both pharmacological intervention and lifestyle intervention. This contributes to better glycemic control compared with r-CGM.

One important factor in judging the efficacy of glucose-lowering therapies is baseline HbA1c, as higher baseline HbA1c is associated with greater reductions in HbA1c [40] if tightly controlled. Regarding the effect of SMBG, glycemic control significantly improved in patients whose baseline HbA1c was ≥ 8% [9]. The impact of baseline HbA1c on the effect of CGM in lowering glucose level has never been evaluated yet. In our meta-analysis, participants from the included studies had varied baseline HbA1c level. Diabetes Center and Joslin Clinic made a clinical recommendation that intensive therapy is need if HbA1c is > 8% [41]. Moreover, significant high risk of microvascular complications was found in subjects with poor glycemic control (HbA1c > 8%) [42]. Therefore, we used HbA1c level of 8% as a cutoff point and the analysis was stratified into three subgroups of baseline HbA1c: <8%, 8 - 10% and > 10%. It was found that CGM was not effective in improving glycemic control regardless of baseline HbA1c. Moreover, subgroup analysis by quality of studies was performed to analyze the effect of study quality on heterogeneity and it was found that quality of studies had no impact on heterogeneity.

The Juvenile Diabetes Research Foundation primary study which compared RT- CGM with SMBG in three age groups (8-14 years, 15 – 24 years and ≥ 25 years) found that CGM was only significantly more effective than SMBG in those aged 25 or more, most likely because older patients used the sensor more frequently since it was found that sensor use was significantly greater in ≥ 25 age group than ≤ 25 age groups (p < 0.001) [26]. The important point related with this finding may be independent of sensor use; older patients are slightly more able than younger patients to interpret data from continuous glucose monitoring to adjust therapy and lifestyle intervention to maintain good glycemic control. To attain better glycemic control in pediatric population, co-operation of parents is demanded. Poor adherence in diabetes management has been recognized as a hindrance to get successful glycemic control in adolescents and children with type 1 diabetes. Moreover, the transition from parental assistance with management of diabetes to patient-only management is often associated with deterioration of glycemic control [43, 44]. According to those findings, RT- CGM can be a useful tool in glycemic control for T1DM pediatric population if the parents and children are carefully educated about the application and benefit of the device. Among 5 RT-CGM studies included in our meta-analysis, only one study indicated that both patients and their parents were educated about managing the device [30], whereas only patients were educated in the remaining four studies. Better results from RT-CGM devices can be expected when not only patients but also their parents are carefully educated about the effect of device on therapy.

Apart from substantial heterogeneity among study results as previously discussed, the limitation of meta-analysis in T1DM pediatrics was the presence of publication bias since only published trials were included. This may lead to omitting appropriate data for the analysis. Other than publication bias, plot asymmetry may be due to small study effect; i.e, the association between estimated intervention effect and the sample size is much different from those might be expected by chance. Consequently, false positive results can be brought about by small studies. In our meta-analysis, we suspected small study effect to be contributed by one study in which the sample size was too small (n = 11) [28]. The effective appropriate frequency and duration of CGM use is also a good point to focus in further studies. In addition, it would be worth analyzing the benefit of CGM device in the populations with nocturnal hypoglycemia, hypoglycemic unawareness where the usefulness of SMBG is limited.

Type 2 Diabetic Adults

It was found that CGM can be a useful tool to reduce HbA1c level in type 2 diabetes adults. Although no heterogeneity was detected (I2 = 0%), there were some variations among the included studies in terms of study quality, types of data presentation (retrospective and real-time), frequency and duration of CGM use, intervention used by the participants. Among four studies included, two studies used retrospective CGM [32, 33] whereas other two studies used real-time CGM [31, 34]. The frequency of CGM use differed greatly from trial to trial, making it difficult to suggest which one is the most appropriate frequency of CGM use. Generally, factors to be considered in determining the frequency of use will include the conditions of patients, i.e. whether their glycemia are poorly controlled or well controlled, and to what extent they are vulnerable to hypoglycemia. Participants were treated with either oral hypoglycemic agents only [33], prandial insulin [34] or both insulin and oral hypoglycemic agents [31, 32].

According to 2006 American Diabetes Association’s recommendation, SMBG is regarded as an essential aspect of diabetes management in insulin-treated patients and a desirable aspect in non-insulin treated patients with diabetes [45]. Although, the effectiveness of SMBG in T2DM patients who do not use insulin had been provided in some meta-analyses, SMBG is not recommended yet. High risk of hypoglycemia can be predisposed by the use of insulin, whereas oral agents particularly metformin, thiazolidinediones, and DPP-4 inhibitors, are not prone to result in hypoglycemia. Moreover, SMBG data or CGM data are generally used to adjust insulin dose. Therefore, in non-insulin treated T2DM, the value of SMBG and CGM previously seems to be limited. In our review, participants in the study by Ehrhardt et al. [34] did not use prandial insulin and participants in the study by Allen et al. [33] did not use any type of insulin at all. These two studies reported the significant reduction in HbA1c in favor of CGM group. This may be because CGM data can provide benefit to lifestyle intervention other than adjusting insulin therapy. In the study by Allen et al. [33], CGM device was used as a counseling tool in evaluating the effect of physical activity on glycemic control and in the study by Ehrhardt et al. [34], patients were allowed to manage their lifestyle based on CGM data. In agreement with the result, CGM use in non-insulin treated T2DM may be beneficial. It is noted that patients enrolled in individual studies had baseline HbA1c > 8%. Thus, CGM devices may be a useful tool in poorly controlled T2DM patients.

When the effectiveness of the r-CGM and RT-CGM compared with SMBG were analyzed separately in T2DM patients, only RT-CGM devices improved glycemic control [mean difference -0.27% (95%CI-0.34 to -0.19)]. One systematic review concluded that RT-CGM may be more effective than r-CGM in type 1 diabetes [39]. However, in type 2 diabetes, whether retrospective CGM or real-time CGM provides superior benefit in lowering HbA1c level compared with SMBG is not known yet. In our meta-analysis, two studies used RT-CGM (Yoo et al. and Ehrhardt et al.) [31, 34] and two studies analyzed data in a retrospective manner (Allen et al. and Cosson et al.) [32, 33]. Both RT-CGM studies and one r-CGM study provided significant difference in lowering HbA1c level by CGM compared with SMBG. However, there is a confounding factor in the study used r-CGM. This r-CGM study was performed to test the role of CGM feedback in physical activity counseling. Therefore, all participants in the study were in physical activity program and these participants were educated about the benefit of physical activity using individual CGM graph or profile at the end of study. As a result, it is difficult to say that lowering HbA1c level was due to CGM device since there was an impact of physical activity on HbA1c level. Therefore, it may be possible that positive effect in our meta-analysis was contributed merely by RT-CGM. Unfortunately, subgroup analysis for quantitative examination of the effect of r-CGM and RT-CGM cannot be performed since the number of studies was too small.

In addition to pharmacological intervention, lifestyle intervention is already accepted as an important aspect in the management of type 2 diabetes. Patients with lifestyle intervention experience significant improvement in blood glucose and lipid profile after one year compared with those without this intervention, supporting the value of lifestyle intervention in T2DM management [46]. Continuous glucose monitoring can provide glucose information promptly to adjust the dosages of medication, and/or to manage the dietary or exercise regimen, which could lead to better lifestyle for diabetes patients. In other words, the glycemic values from CGM devices allow care provider or patient to encounter the noticeable effects of meals and exercise on glycemic level and to manage lifestyle skills that results in better glycemic control. Moreover, from the glycemic values displayed by CGM devices, patients can be easily educated about the effects of their medication and lifestyle on diabetes. A study testing the effects of counseling intervention of physical activity and using continuous glucose monitoring system feedback suggested that counseling T2DM patients with CGM feedback may improve physical activity level and, as a result, diabetes-related complications could be reduced [33]. Therefore, with the help of continuous glucose monitoring, not only pharmacological intervention can be adjusted, but non-pharmacological intervention can also be promoted.

There were some limitations in the meta-analysis of type 2 diabetes. We did not search for unpublished trials. Thus there was a possibility of missing relevant data. However, no publication bias was detected. The average baseline HbA1c of participants from all studies was above 8%. Therefore, the routine use of CGM in patients with uncontrolled T2DM may provide benefit and deserves consideration. More randomized trials with adequate sample sized are needed for CGM use in type 2 diabetes population. In addition, further studies are suggested to focus on the frequency and duration of RT-CGM use to achieve the legacy effects.

Conclusion

Type 1 diabetic pediatrics

The available evidence from this meta-analysis suggests that CGM use was no more effective than SMBG in reducing HbA1c in type 1 diabetic pediatrics. However, the results of subgroup analyses showed that RT-CGM can be more effective than SMBG in reducing HbA1c. In contrast, r-CGM was no better than SMBG in T1DM pediatric population. Therefore, RT-CGM devices can be an effective tool in Type 1 diabetic pediatrics population.

Type 2 diabetic adults

Our results suggest that the effect of CGM use in lowering HbA1c level was superior to SMBG for type 2 diabetes adults. Since the average baseline HbA1c of participants from all studies was above 8%, CGM device can be an effective tool in patients with uncontrolled T2DM. However, the comprehensive randomized controlled trial need to be established in order to address the long term benefit of CGM.

Abbreviations

ADA:

American diabetes association

AUC:

Area under the curve

BG:

Blood glucose

CGM:

Continuous glucose monitoring

HbA1c:

Hemoglobin A1c

IDDM:

Insulin dependent diabetes mellitus

NIDDM:

Non-insulin dependent diabetes mellitus

OHA:

Oral hypoglycemic agents

RCT:

Randomized controlled trial

r-CGM:

retrospective CGM

RT-CGM:

Real-time CGM

SMBG:

Self monitoring of blood glucose

T1DM:

Type 1 Diabetes Mellitus

T2DM:

Type 2 Diabetes mellitus

UKPDS:

UK Prospective diabetes study

95% CI:

95% Confidence interval.

References

  1. Whiting DR, Guariguata L, Weil C, Shaw J: IDF diabetes atlas: global estimates of the prevalence of diabetes for 2011 and 2030. Diabetes Res Clin Pract. 2011, 94 (3): 311-321. 10.1016/j.diabres.2011.10.029.

    Article  PubMed  Google Scholar 

  2. Wild S, Roglic G, Green A, Sicree R, King H: Global prevalence of diabetes: estimates for the year 2000 and projections for 2030. Diabetes Care. 2004, 27 (5): 1047-1053. 10.2337/diacare.27.5.1047.

    Article  PubMed  Google Scholar 

  3. International Diabetes Federation. Facts & figures: prevalence. 2006,http://www.idf.org/diabetesatlas/5e/what-is-diabetes,

  4. Hoerger TJ, Segel JE, Gregg EW, Saaddine JB: Is glycemic control improving in U.S. adults?. Diabetes Care. 2008, 31: 81-86. 10.2337/dc08-1303.

    Article  PubMed  Google Scholar 

  5. UK Prospective Diabetes Study (UKPDS) Group: Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 Diabetes (UKPDS 33). Lancet. 1998, 652: 837-853.

    Google Scholar 

  6. The Diabetes Control Complications Research Group: The effect of intensive treatment of diabetes on the development and progression of long term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993, 329: 977-986.

    Article  Google Scholar 

  7. Gold EA, Mac Leod KM, Frier B: Frequency of severe hypoglycaemia in patients with type 1 diabetes with impaired awareness of hypoglycaemia. Diabetes Care. 1994, 17: 697-703. 10.2337/diacare.17.7.697.

    Article  CAS  PubMed  Google Scholar 

  8. Sarol JN, Nicodemus NA, Tan KM, Grava MB: Self-monitoring of blood glucose as part of a multi-component therapy among non-insulin requiring type 2 diabetes patients: a meta-analysis (1996-2004). Curr Med Res Opin. 2005, 21 (2): 173-184. 10.1185/030079904X20286.

    Article  PubMed  Google Scholar 

  9. Poolsup N, Suksomboon N, Rattanasookchit S: Meta-analysis of the benefits of self-monitoring of blood glucose on glycemic control in type 2 diabetes patients: An update. Diabetes Techno Ther. 2009, 11 (12): 775-782. 10.1089/dia.2009.0091.

    Article  CAS  Google Scholar 

  10. Weber KK, Lohmann T, Busch K, Donati-Hirsch I, Riel R: High frequency of unrecognized hypoglycaemias in patients with type 2 diabetes is discovered by continuous glucose monitoring. Exp Clin Endocrinol Diabetes. 2007, 115: 491-494. 10.1055/s-2007-984452.

    Article  CAS  PubMed  Google Scholar 

  11. Boland E, Monsod T, Delucia M, et al: Limitations of conventional methods of self-monitoring of blood glucose: lessons learned from 3 days of continuous glucose sensing in pediatric patients with type 1 diabetes. Diabetes Care. 2001, 24 (11): 1858-1862. 10.2337/diacare.24.11.1858.

    Article  CAS  PubMed  Google Scholar 

  12. MacGowan K, Thomas W, Moran A: Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes. Diabetes Care. 2002, 25 (9): 1499-1503. 10.2337/diacare.25.9.1499.

    Article  Google Scholar 

  13. Kerr D, Fayers K: Continuous real-time glucose monitoring systems: time for a closer look. Pract Diab Int. 2008, 25 (1): 37-41. 10.1002/pdi.1197.

    Article  Google Scholar 

  14. Bergenstal RM, Tamborlane WV, Ahmann A, et al: Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Eng J Med. 2010, 363: 311-320. 10.1056/NEJMoa1002853.

    Article  CAS  Google Scholar 

  15. Battelino T, Congret I, Olsen B, et al: The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial. Diabetologia. 2012, 55 (12): 3155-3162. 10.1007/s00125-012-2708-9.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  16. Golicki DT, Golicka D, Groele L, Pankowska E: Continuous glucose monitoring system in children with type 1 diabetes mellitus: a systematic review and meta-analysis. Diabetologia. 2008, 51: 233-240. 10.1007/s00125-007-0884-9.

    Article  CAS  PubMed  Google Scholar 

  17. Gandhi GY, Kovalaske M, Kudva Y, et al: Efficacy of continuous glucose monitoring in improving glycemic control and reducing hypoglycemia: A systematic Review and meta-analysis of randomized trial. J Diabetes Sci Technol. 2011, 5 (4): 952-963.

    Article  PubMed Central  PubMed  Google Scholar 

  18. Furlan AD, Pennick V, Bombardier C, Tulder MV: 2009 updated method guidelines for systematic reviews in the cochrane back review group. Spine. 2009, 34 (18): 1929-1941. 10.1097/BRS.0b013e3181b1c99f.

    Article  PubMed  Google Scholar 

  19. Jadad AR, Moore RA, Carroll D, Jenkinson C, Reynolds DJ, Gavaghan DJ, et al: Assessing the quality of reports of randomized clinical trials: is blinding necessary?. Control Clin Trials. 1996, 17 (1): 1-12. 10.1016/0197-2456(95)00134-4.

    Article  CAS  PubMed  Google Scholar 

  20. Verhagen AP, de Vet HC, de Bie RA, Kessels AG, Bouter LM, Knipschild PG: The Delphi list: a criteria list for quality assessment of randomized clinical trials for conducting systematic reviews developed by Delphi consensus. J Clin Epidemiol. 1998, 51: 1235-1241. 10.1016/S0895-4356(98)00131-0.

    Article  CAS  PubMed  Google Scholar 

  21. Deeks JJ, Altman DG, Bradburn MJ: Statistical methods for examining heterogeneity and combining results from several studies in meta-analysis. Systematic reviews in health care: meta-analysis in context. Edited by: Egger M, Smith GD, Altman DG. 2001, United Kingdom: BMJ publishing group, 289-290. 2

    Google Scholar 

  22. Egger M, Davey SG, Schneider M, Minder C: Bias in meta-analysis detected by a simple, graphical test. BMJ. 1997, 315: 629-634. 10.1136/bmj.315.7109.629.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  23. Deiss D, Hartmann R, Schmidt J, Kordonouri O: Results of a randomized controlled cross-over trial on the effect of continuous subcutaneous glucose monitoring (CGMS) on glycemic control in children and adolescents with type 1 diabetes. Exp Clin Endocrinol Diabetes. 2006, 114: 63-67. 10.1055/s-2006-923887.

    Article  CAS  PubMed  Google Scholar 

  24. Lagarde WH, Barrows FP, Davenport ML, Kang M, et al: Continuous subtaneous glucose monitoirng in children with type 1 diabetes mellitus: a single-blind, randomized, controlled trial. Paediatric Diabetes. 2006, 7: 159-164. 10.1111/j.1399-543X.2006.00162.x.

    Article  Google Scholar 

  25. Yates K, Milton AH, Dear K, Ambler G: Continuous glucose monitoring-guided insulin adjustment in children and adolescents on near-physiological insulin regimens. Diabetes Care. 2006, 29 (7): 1512-1517. 10.2337/dc05-2315.

    Article  CAS  PubMed  Google Scholar 

  26. The juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group: Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med. 2008, 359: 1464-1476.

    Article  Google Scholar 

  27. Kordonouri O, Pankowska E, Rami B, et al: Sensor-augmented pump therapy from the diagnosis of childhood type 1 diabetes: results of the Paediatric Onset Study (ONSET) after 12 months of treatment. Diabetologia. 2010, 53: 2487-2495. 10.1007/s00125-010-1878-6.

    Article  CAS  PubMed  Google Scholar 

  28. Chase HP, Kim LM, Owen SL, Mackenzie TA, Klingersmith GJ, Murtfeld R, Garg SK: Continuous subcutaneous glucose monitoring in children with type 1 diabetes. Pediatrics. 2001, 107: 222-226. 10.1542/peds.107.2.222.

    Article  CAS  PubMed  Google Scholar 

  29. Bukara-Radujkovic G, Zdravkovic D, Lakic S: Short-term use of continuous glucose monitoring system adds to glycemic control in young type 1 diabetes mellitus patients in the long run: a clinical trial. Vojnosanit Pregl. 2011, 68 (8): 650-654. 10.2298/VSP1108650B.

    Article  PubMed  Google Scholar 

  30. Mauras N, Beck R, Xing D, et al: A randomized clinical trial to assess the efficacy and safety of real-time continuous glucose monitoring in the management of type 1 diabetes in young children aged 4 to < 10 years. Diabetes Care. 2012, 35: 204-210. 10.2337/dc11-1746.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  31. Yoo HJ, An HG, Park SY, et al: Use of a real time continuouse glucose monitoring system as a motivational device for poorly controlled type 2 diabetes. Diabetes Res Clin Pract. 2008, 82: 73-79. 10.1016/j.diabres.2008.06.015.

    Article  CAS  PubMed  Google Scholar 

  32. Cosson E, Hamo-Tchatchouang E, Dufaitre-Patouraux L, et al: Multicentre, randomized, controlled study of the impact of continuous sub-cutaneous glucose monitoring (GlucoDay) on glycaemic control in type 1 and type 2 diabetic patients. Diabetes Metab. 2009, 35 (4): 312-318. 10.1016/j.diabet.2009.02.006.

    Article  CAS  PubMed  Google Scholar 

  33. Allen NA, Fain JA, Braun B, Chipkin SR: Continuous glucose monitoring improves physical activity behaviors of individuals with type 2 diabetes: a randomized clinical trial. Diabetes Res Clin Pract. 2008, 80: 371-379. 10.1016/j.diabres.2008.01.006.

    Article  PubMed Central  PubMed  Google Scholar 

  34. Ehrhardt NM, Chellappa M, Walker MS, Fonda SJ, Vigersky RA: The effect of real-time continuous glucose monitoring on glycemic control in patients with type 2 diabetes mellitus. J Diabetes Sci Technol. 2011, 5 (3): 668-675.

    Article  PubMed Central  PubMed  Google Scholar 

  35. Vigersky RA, Fonda SJ, Chellappa M, Walker MS, Ehrhardt NM: Short- and Long- term effects of real-time continuous glucose monitoring in patients with type 2 diabetes. Diabetes Care. 2012, 35: 32-38. 10.2337/dc11-1438.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  36. American Academy of Paediatric: Definition of age group terminology.http://www.aap.org/en-us/Pages/Default.aspx,

  37. Ludvigsson J, Hanas R: Continuous subcutaneous glucose monitoring improved metabolic control in pediatric patients with type 1 diabetes: a controlled crossover study. Pediatrics. 2003, 111 (5 pt 1): 933-938.

    Article  PubMed  Google Scholar 

  38. Wojciechowski P, Rys P, Lipowska A, et al: Efficacy and safety comparison of continuous glucose monitoring and self-monitoring of blood glucose in type 1 diabetes. Pol Arch Med Wewn. 2011, 121 (10): 334-340.

    Google Scholar 

  39. Langendam M, Luijf YM, Hooft L, DeVries JH, Mudde AH, Scholten RJPM: Continuous glucose monitoring systems for type 1 diabetes mellitus. Cochrane Database Syst Rev. 2012, 2: 1-135.

    Google Scholar 

  40. DeFronzo RA, Stonehouse AH, Han J, Wintle ME: Relationship of baseline HbA1c and efficacy of current glucose-lowering therapies: a meta-analysis of randomized clinical trials. Diabet Med. 2010, 27 (3): 309-317. 10.1111/j.1464-5491.2010.02941.x.

    Article  CAS  PubMed  Google Scholar 

  41. Joslin Diabetes Center & Joslin Clinic: Clinical guidelines for adults with diabetes. [online]. http://www.joslin.org/info/joslin-clinical-guidelines.html

  42. Krolewski AS, Laffel LM, Krolewski M, Quinn M, Warram JH: Glycosylated hemoglobin and the risk of microalbuminuria in patients with insulin-dependent diabetes mellitus. N Engl J Med. 1995, 332 (19): 1251-1255. 10.1056/NEJM199505113321902.

    Article  CAS  PubMed  Google Scholar 

  43. Anderson BJ, Vangsness L, Connell A, Butler D, Goebel-Fabbri A, Laffel LMB: Family conflict, adherence, and glycaemic control in youth with short duration Type 1 diabetes. Diabet Med. 2002, 19: 635-642. 10.1046/j.1464-5491.2002.00752.x.

    Article  PubMed  Google Scholar 

  44. Insabella G, Grey M, Knafl G, Tamborlane W: The transition to young adulthood in youth with type 1 diabetes on intensive treatment. Pediatr Diabetes. 2007, 8: 228-234. 10.1111/j.1399-5448.2007.00266.x.

    Article  PubMed  Google Scholar 

  45. American Diabetes Association: Standards of medical care in diabetes—2006. Diabetes Care. 2006, 29 (Suppl 1): S4-S42.

    Google Scholar 

  46. Mitra A, Dewanjee D, Dey B: Mechanistic studies of lifestyle interventions in type 2 diabetes. World J Diabetes. 2012, 3 (12): 201-207. 10.4239/wjd.v3.i12.201.

    Article  PubMed Central  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Naeti Suksomboon.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contribution

NS contributed to conception, design, analysis and interpretation of data and drafted the manuscript. NP and AMK contributed to design, analysis and interpretation of data and drafted the manuscript. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Poolsup, N., Suksomboon, N. & Kyaw, A.M. Systematic review and meta-analysis of the effectiveness of continuous glucose monitoring (CGM) on glucose control in diabetes. Diabetol Metab Syndr 5, 39 (2013). https://doi.org/10.1186/1758-5996-5-39

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1758-5996-5-39

Keywords